Calibrate AfiLabs
In addition to routine cleaning, AfiLab requires regular calibration to ensure accurate milk measurements (we recommend at least every second month and after a major change in cows' feed rations).
The following types of calibration are required:
-
Calibrate AfiLabs - Bulk Tank Method – automatic calibration based on milk shipment measurements (Fat/Protein/Lactose) received from the milk processing plant.
-
Calibrate AfiLab - Milk Test Method – Calibration based on Milk Test Results performed monthly. Several methods are available to import the data or enter it manually.
-
Periodic Calibration by a technician.
Note: If no calibration was performed for 6 months, the system displays an alert notification. Two weeks later the system calibrates AfiLab to the factory default values.